7. Important information
Electromagnetic immunity
The Eonis is intended for use in the electromagnetic environment specied below. The customer or the
user of the Eonis should assure that it is used in such an environment.
Immunity test IEC 60601
Test levels
Compliance level Electromagnetic
environment
guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
± 6kV contact
±8kVair
± 6kV contact
±8kVair
Floors should be wood,
concrete or ceramic tile.
If oors are covered with
synthetic material, the
relative humidity should
be at least 30%
Electrical fast
transient/burst
IEC 61000-4-4
± 2kV for power supply
lines
± 1kV for input/ output
lines
± 2kV for power supply
lines
± 1kV for input/ output
lines
Mains power quality
should be that of a typical
commercial or hospital
environment
Surge
IEC61000-4-5
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
± 1 kV line(s) to line(s)
±2kVline(s)toearth
Mains power quality
should be that of a typical
commercial or hospital
environment
Voltage dips, short
interruptions and voltage
variations on power
supply input lines
IEC 61000-4-11
<5%U
T
1
(> 95% dip in
U
T
) for 0.5 cycle
40% U
T
(60%dipinU
T
)
for 5 cycles
70% U
T
(30%dipinU
T
)
for 25 cycles
<5%U
T
(>95% dip in
U
T
) for 5s
<5%U
T
(> 95% dip in
U
T
) for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
<5%U
T
(>95% dip in
U
T
)for5s
Mains power quality
should by that of a
typical commercial or
hospital environment. If
the user of the Eonis
requires continued
operation during power
mains interruptions, it is
recommended that the
Eonis be powered from
an uninterruptible power
supply or a battery.
Power frequency (50/60
Hz) magnetic eld
IEC 61000-4-8
3A/m
Not applicable
2
Power frequency
magnetic elds should
be at levels characteristic
of a typical location in
a typical commercial or
hospital environment.
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3Vrms
150 kHz to 80 MHz
3V/m
80 MHz to 2.5 GHz
3V
3V/m
Portable and mobile
RF communications
equipment should be
used no closer to
any part of the Eonis,
including cables, than
the recommended
separation distance
calculated from the
equation applicable
to the frequency
of the transmitter.
Recommended
separation distance
d=1.2P
d=1.2P 80 MHz to 800
MHz
1. is the a.c. mains voltage prior to application of the test level.
2. Eonis doesn’t contain susceptible components to magnetic fields
K5903081 EONIS 20/01/2014 27